Heartbeat detection

ECG-based sleep staging heavily relies on heartrate variability. Therefore, a reliable and efficient heartbeat detector is essential. SleepECG provides a detector based on the approach described by Pan & Tompkins (1985). We outsourced performance-critical code to a C extension, which makes the detector substantially faster than other existing Python implementations. However, we also provide Numba and pure Python backends (the Numba backend is almost as fast, whereas the pure Python implementation is much slower).

Usage

The function detect_heartbeats() finds heartbeats in an unfiltered ECG signal ecg with sampling frequency fs (in Hz). It returns the indices of all detected heartbeats. A complete example including visualization and performance evaluation is available in examples/heartbeat_detection.py.

from sleepecg import detect_heartbeats

beats = detect_heartbeats(ecg, fs)

For best results, we recommend a sampling frequency of at least 100 Hz. Furthermore, the algorithm will return similar results regardless of the scaling of the data.

Examples

Let's detect heartbeats in a short electrocardiogram:

import numpy as np
from sleepecg import detect_heartbeats, get_toy_ecg

ecg, fs = get_toy_ecg()  # 5 min of ECG data at 360 Hz
beats = detect_heartbeats(ecg, fs)

Next, we calculate RR intervals (in milliseconds):

rri = 1000 * np.diff(beats) / fs

Now let's add some random noise to the electrocardiogram and compare the heartbeat detection with and without noise:

import numpy as np
from sleepecg import compare_heartbeats

np.random.seed(42)
ecg_noisy = ecg + np.random.rand(ecg.size)
beats_noisy = detect_heartbeats(ecg_noisy, fs)
tp, fp, fn = compare_heartbeats(beats_noisy, beats)

Benchmarks

All code used for performance evaluation is available in examples/benchmark/. The used package versions are listed in requirements-benchmark.txt.

We evaluated detector runtime using slices of different lengths from LTDB records with at least 20 hours duration. The C backend of our detector is by far the fastest implementation among all tested packages (note that the y-axis is logarithmically scaled). Runtime evaluation was performed on an Intel® Xeon® Prozessor E5-2440 v2 with 32 GiB RAM. No parallelization was used.

LTDB runtimes

We also evaluated detection performance on all MITDB records. We defined a successful detection if it was within 100 ms (36 samples) of the corresponding annotation (using a tolerance here is necessary because annotations usually do not coincide with the exact R peak locations). In terms of recall, precision, and F1 score, our detector is among the best heartbeat detectors available.

MITDB metrics

For analysis of heartrate variability, detecting the exact location of heartbeats is essential. As a measure of how accurate a detector is, we computed Pearson's correlation coefficient between resampled RRI time series deduced from annotated and detected beat locations from all GUDB records. Our implementation detects peaks in the bandpass-filtered ECG signal, so it produces stable RRI time series without any post-processing.

GUDB pearson correlation